The first time an ultrasonic cleaning solution underperforms, the chemistry takes the blame. Yet the more likely cause of mid-shift failures is concentration drift, a slow change that make-up water and part drag-out introduce without warning. This post treats concentration as a controlled variable rather than a pour-and-hope habit, and it shows you how to keep it on spec.
You will learn why cleaners seem to stop working mid-shift, what to have in place before you dose anything, and how to set a titrated baseline for each tank. We cover choosing between titration and metering with a fixed check interval, correcting for make-up water chemistry, accounting for drag-out, and logging every reading so results repeat shift to shift and your validation file stays defensible. Closing tips address over-dosing, top-up versus replace, and matching chemistry to the application.
Kleentek supports that discipline with matched ultrasonic cleaning chemicals and machines built to over 20 W/L power density, plus cleaning trials and training. The company holds ISO 9001:2015 certification and a 12-month warranty.
Why Cleaners ‘Stop Working’ Mid-Shift: It Is Usually Concentration Drift
When parts emerge filmed, spotted or inconsistent, the working concentration of your ultrasonic cleaning solution has usually drifted outside its specified range. Before blaming the chemistry, check the bath.
Across a single shift, four routine drivers move that number: evaporative losses, make-up water additions, chemical carried out on parts and baskets (drag-out), and soil loading from the workload. Heavy contamination loads, such as those on mining and heavy engineering components, accelerate all four.
Concentration is a process variable with a specification, exactly like temperature, dwell time and power density. Even at Kleentek’s rated power density, cavitation cannot compensate for a bath outside its specified concentration.
Diagnose in order: verify concentration, then bath temperature, then mechanical and ultrasonic function, then whether the chemistry matches the substrate and soil. In our experience on production lines, most ‘the cleaner stopped working’ complaints resolve at step one. The same logic applies to multi-stage lines, where rinse stages drift too.
Prerequisites: What You Need in Place Before You Dose Anything
Once you accept that drift is routine, the next question is what the bath needs before any dose goes in. Without these five items, concentration control is guesswork.
- Matched ultrasonic cleaning solution with a technical data sheet. The TDS must state the working concentration range, the recommended titration or metering method, and make-up instructions. A container with no data sheet is not a controlled process input.
- A measurement method. A titration kit (burette, indicator, sample vessels, pipettes) suits periodic verification. If readings are needed every shift, use a calibrated bench or inline meter.
- Water-quality data for your make-up water. Hardness, conductivity and dissolved solids, from your water supplier or an on-site test. This is the figure most sites are missing when readings stop making sense.
- A log template. Fields for date, shift, tank ID, bath volume, reading, water source, corrective action, chemical added and operator initials.
- Documented acceptance limits and check intervals. Source these from the chemistry supplier’s technical data sheet and your own validation requirements, not set on the fly by whoever is on shift.
Optional but useful: Kleentek cleaning trials establish a baseline recipe and starting concentration for your specific parts and soils before formal limits are written into the work instruction.
Bath volume, tank configuration and power density sit in the same specification exercise. See our guide to sizing and specifying a large industrial ultrasonic cleaner for tanks from 500 to 20,000 litres.
Step 1: Establish a Titrated Baseline for Each Tank
With those inputs in place, the first control action is to fix a known starting point.
After commissioning a fresh charge, circulate or degas the bath as the chemistry supplier directs, bring it to working temperature, then draw a sample from a fixed point in the tank, not from the surface or beside a drain. Titrate or meter the sample and confirm the reading sits inside the specified range for your measured bath volume. Record that figure as the baseline for that tank.
Set a separate baseline for every chemistry-bearing stage. A cleaning tank, a rinse tank and a downstream chemistry stage drift at different rates and for different reasons, so one line-wide figure proves nothing.
Note the reference conditions beside each baseline: part family, basket type, load size and cycle. These govern how much solution leaves the tank per cycle, so a baseline without them is not comparable. Where those conditions are not yet characterised for a part and soil, a structured ultrasonic cleaning trial is the practical route.
Keep the baseline sheet with the validation file. That way a new operator, a new shift or a new week starts from exactly the same documented point.
Step 2: Choose Titration or Metering, Then Fix Your Check Interval
With baselines documented, decide how you will measure against them. Titration gives a direct measurement of active chemical in the bath. It suits smaller tanks, low-throughput lines, and periodic verification of any metered stage.
Metering is faster: conductivity or refractive index, depending on the chemistry. It suits high-throughput industrial lines where a reading is needed every shift or batch. Whichever you use, verify the meter against titration at defined intervals. Probes drift with soil loading, temperature and chemical change, and a drifted probe returns confident wrong numbers.
Set the check interval from the chemistry supplier’s technical data sheet and your validation requirements. Once per shift is a reasonable starting point in practice; your chemistry supplier’s TDS and validation requirements govern the actual interval. Heavy soiling or large drag-out may justify a mid-shift check. High-consequence work, such as aerospace and defence cleaning, typically calls for tighter, documented intervals as specified by the relevant sector standard or customer requirement.
Sample consistently. Same tank location, same point in the cycle, same temperature, every time. Otherwise readings are not comparable between shifts, and the trend becomes noise.
Write a response plan. Below range: top up and re-check. Above range: dilute or replace as the supplier directs. Put both actions, and the acceptance limits, in the work instruction. A reading with no defined action is only a record.
Step 3: Correct for Make-Up Water Chemistry
With sampling point and interval fixed, the next variable is what you pour in. Top-up water is not neutral: hardness, conductivity and dissolved solids add ions that shift readings and can interact with the detergent chemistry. Even tap and distilled water carry minerals and organics to some degree, and hard water can react with the chemistry and leave white residue on parts.
If your supply varies (town water, bore, recycled or tankered), document which approach you take. Treat it by softening, reverse osmosis or deionisation, or correct for the supply chemistry using your chemical supplier’s guidance. The same discipline applies to chemistry selection, as our guide to choosing an ultrasonic cleaning fluid for fuel injectors explains.
Record the water source against every top-up entry. A step change in readings can then be traced to a source change rather than triggering an unnecessary chemical replacement.
Never top up to a volume mark without a reading. Volume-based top-up assumes water chemistry is constant, and that is exactly the assumption that fails when a supply changes.
When a reading looks wrong but the bath behaves normally, suspect the water before you suspect the ultrasonic cleaner solution. Where water quality varies materially, state it as a known limitation in the validation file, so an auditor sees the control you applied rather than an unexplained variation. Top-up corrects what evaporates; what leaves on parts is next.
Step 4: Account for Drag-Out, Not Just Evaporation
Drag-out is often the larger source of drift across a busy shift. Every part and basket leaving a tank carries ultrasonic cleaning solution with it, and on complex geometry and busy cycles that carry-over can strip more active chemical over a shift than evaporation does.
Quantify it rather than guess it. Kleentek cleaning trials, combined with your own measurements across a defined load, give a real carry-over figure for your parts. Dose to that measured loss instead of topping up to a fixed volume, then re-check concentration after the busiest part of the shift, when carry-over peaks.
Carry-over shifts with variables you control: dwell time, drain time, basket design, part orientation and transfer speed. Document each in the work instruction, not operator preference.
Rinse stages drift too. Solution dragged from the cleaning tank dilutes the rinse bank and moves chemistry downstream, so rinse readings matter as much as cleaning tank readings. Where rinse chemistry sits inside a validated process, as in pharmaceutical and food production, that reading belongs in the log alongside the cleaning tank figure.
Step 5: Log Every Reading So Results Repeat
Drag-out tells you how much chemistry leaves the tank. The log tells you what that loss does to the bath over time.
At every check, record the reading, top-up volume, water source, corrective action, chemical added and operator initials. The trend across a shift matters more than any single number.
A reading drifting consistently one way signals that something has changed, such as a leak, an over-dose, a different part mix or a new water source, before parts fail. Keep logs in the validation file, anchored to your quality system. ISO guidance treats retained documented information as evidence that planned processes were followed, and Kleentek operates under an ISO 9001:2015 quality management system, a framework buyers in regulated sectors expect.
On PLC and HMI controlled lines, digital logging helps reduce transcription errors and keeps a timestamped record that survives shift handovers and staff changes. If those records sit on networked systems, manage access and data handling as set out in Kleentek’s privacy policy.
Periodically, for many lines, monthly is practical, reconcile chemical consumption against logged additions. Unaccounted chemical is usually undocumented top-up, and it undermines what the logs prove.
Tips: Over-Dosing, Top-Up Versus Replace, and Matching Chemistry to the Application
Logs only prove control if the dose behind them is correct. Over-dosing feels like insurance, but it lifts chemical cost, leaves residues on parts and can attack the wrong substrate. Dose to specification, then measure to confirm the bath landed where you intended.
Replace rather than top up on a schedule your chemistry supplier and soiling rate justify. A bath at the correct concentration but carrying a heavy soil load is still a failed bath, and a reading will not tell you that. Filtration controls soil, not concentration; heavy soiling can mask drift and foul a conductivity probe, so clean and verify sensors routinely.
Match chemistry to the application before dose. Substrate, soil type, required cleanliness and the downstream process drive the choice. The right ultrasonic cleaning solution for an aluminium component and the right cleaning solution for an ultrasonic cleaner running hardened steel differ in pH and inhibitor chemistry. The best ultrasonic cleaning solution for your line is specified for your substrate, soil and process, not the cheapest container on the shelf. The correct chemistry match significantly improves the likelihood of a cycle that works first time.
Never dose by colour, smell or feel. Those cues are not measurements and cannot be written into a work instruction or validation file.
Undersized or bargain-hardware units cannot be rescued by chemistry. Machines built to a price, without matched chemicals, filtration or local support, leave you guessing whether the tank, the reading or the solution is at fault. A properly specified tank with quality generators and matched chemistry gives stable conditions to control against. For trans-Tasman buyers, this guide to industrial ultrasonic cleaning equipment in New Zealand covers capability, chemistry and support criteria.
Conclusion: Make Concentration a Controlled Variable
Treating concentration with the same discipline as temperature, dwell time and power density is what turns cleaning results from variable to repeatable. Done consistently, that routine delivers results from the first shift of the week to the last, and makes your validation file defensible when a customer or auditor asks how you know the process is in control. The routine, baseline, measure, correct and log, delivers that consistency shift to shift.
Where you need to establish the starting point for a specific part, soil and chemistry combination, Kleentek offers cleaning trials, matched ultrasonic cleaning chemicals, and operator training and support, backed by a 12 month warranty and an ISO 9001:2015 quality management system. Talk to Kleentek about industrial ultrasonic cleaning equipment engineered and supported across Australia and New Zealand.